89
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
See the Model 333 HART
®
Tri-Loop HART-to-Analog Signal Converter Reference Manual
(Document No. 00809-0100-4754) for further information on how to install and configure the
Tri-Loop.
To turn off the Burst Mode
To turn off the Burst Mode, use one of the following options:
The RCT program (requires that the receive and transfer buffers for the selected COM
port are adjusted)
The Rosemount Burst Mode Switch software
A Field Communicator
The AMS
®
Device Manager software
To set the COM port buffers
To be able to communicate with the Rosemount 3300 Series in Burst Mode, the receive and
transfer buffers need to be adjusted as written below.
If you have Windows XP
1. In the MS Windows
™
Control Panel open the System option.
2. Choose the Hardware tab and select the Device Manager button.
3. Expand the Ports node in the tree view.
4. Click the right mouse button on the selected COM port and choose Properties.
5. Select the Port Settings tab and click the Advanced button.
6. Drag the Receive Buffer and Transfer Buffer slides to 1.
7. Select the OK button.
8. Reboot the computer.
If you have Windows 7, Windows 8, or Windows 8.1
1. In the Start menu, select Control Panel > System and Security > System > Device
Manager.
2. Double-click Ports (COM & LPT).
3. Right-click on the selected COM port and select Properties.
4. Select the Port Settings tab and then select the Advanced button.
5. Drag the Receive Buffer and Transmit Buffer to 1.
6. Click OK.
7. Restart the computer.
Comentarios a estos manuales