
69
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
4.4 Basic configuration
This section describes the various HART commands used to configure the Rosemount 3300
Series Transmitters for level measurements. The transmitter outputs a 4-20 mA signal
proportional to the primary variable. Three additional variables are available through the HART
signal.
Table 4-1. Device Revision 2, HART Communicator Fast Key Sequence
4.4.1 Transmitter variables
You can assign up to four transmitter variables. Typically, the Primary Variable (PV) is configured
to be Aggregate Level, Interface Level, or Volume.
For Rosemount 3301, the PV is typically set to be Level. If the transmitter is in the Immerse
Probe mode (see Measurement mode on page 71), the PV is normally set to Interface Level.
For Rosemount 3302, the PV is typically set to Interface Level, but Level and other options can
also be used.
Function Fast Key Sequence
Transmitter variables
(1)
(1) For volume measurement, select the Volume option.
1, 1, 1, 1
Measurement units 1, 3, 1
Reference gauge height 1, 3, 2, 1
Probe length 1, 3, 2, 2
Probe type 1, 3, 2, 3
Product dielectric 1, 3, 3, 3
Vapor dielectric 1, 3, 3, 2
Measurement mode 1, 3, 3, 4
Probe angle 1, 3, 2, 4
Maximum upper product thickness 1, 4, 5, 2
Damping 1, 3, 5
Display panel 1, 4, 2
4 and 20 mA points 1, 3, 4, 3
Volume units 1, 3, 1, 2
Tank t ype 1, 4, 3, 1
Tank dimensions 1, 4, 3, 2-3
Strapping table 1, 4, 3, 4
HART Comm
1, 1, 1, 1
Comentarios a estos manuales