Section 3: Installation
January 2015
Installation
3.5.8 Anchoring
In turbulent tanks, it may be necessary to fix the probe. Depending on the probe type, different
methods can be used to guide the probe to the tank bottom. This may be needed to prevent the
probe from hitting the tank wall or other objects in the tank, as well as preventing a probe from
breaking.
Flexible twin/single lead probe with weight
and ring
A ring (customer supplied) can be attached to the
weight in a threaded (M8x14) hole at the end of
the weight. Attach the ring to a suitable
anchoring point.
Flexible twin/single lead probe with weight
and magnet
A magnet (customer supplied) can be fastened in
a threaded (M8x14) hole at the end of the weight.
The probe can then be guided by placing a
suitable metal plate beneath the magnet.
Coaxial probe fixed to the tank wall
The coaxial probe can be guided to the tank wall
by fixtures fastened to the tank wall. Fixtures are
customer supplied. Make sure the probe can
move freely due to thermal expansion without
getting stuck in the fixture.
Weight with
internal threads
M8x14
Ring
Magnet
1.1 in. (28 mm)