30
Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
3.5.2 Tank connection with loose flange (“plate design”)
Figure 3-12. Tank Connection with Loose Flange (“Plate Design”)
The transmitter is delivered with head, flange, and probe assembled into one unit. If, for some
reason, these parts are disassembled, mount the transmitter as described below:
1. Place a gasket on top of the tank flange.
2. Mount the flange on the probe and tighten the flange nut.
3. Mount the transmitter head.
4. Lower the transmitter and probe with flange into the tank.
5. Tighten the bolts.
6. Loosen the nut that connects the transmitter housing to the probe slightly.
7. Rotate the transmitter housing so that the cable entries/display face the desired
direction.
8. Tighten the nut.
Note
PTFE covered probes must be handled carefully to prevent damage to the coating.
E
H
G
D
C
A
B
F
A Flange nut
B Bolts
C Probe
D Tank flange
E Transmitter head
F Nut
G Flange
H Gasket
Comentarios a estos manuales